Model 4510
Conductivity/TDS Meter
Operating Manual
451 050/REV A/03-03
Vista de página 0
1 2 3 4 5 6 ... 22 23

Resumo do Conteúdo

Página 1 - Operating Manual

Model 4510Conductivity/TDS MeterOperating Manual451 050/REV A/03-03

Página 2

3.2 Instrument Set-UpThe following section details the set-up modes available to the user.These are: Cell constant, Coefficient, Reference Temperature

Página 3

3.2.2 Temperature Coefficient set upThe display will show:The secondary display will show COEFF and will then scroll TEMPERATURE COEFFICIENT after10 s

Página 4

3.2.4 TDS EC RatioThe display will show:The secondary display will show TDS EC and will then scroll TDS EC RATIO after 10 seconds.Symbol will fl

Página 5 - 451 050/REV A/03-032

3.2.6 Baud RateFor full details relating to serial protocol refer to Section 6 of this manual.The display will show:The secondary display will show BA

Página 6

3.5 Calibration with Standard SolutionCalibration of the unit and cell with standard solutions will only be necessary if:1) The cell constant is unkno

Página 7

3.6 Sample MeasurementGeneralConductivity is a temperature dependent measurement. All substances have a conductivitycoefficient which varies from 1% p

Página 8 - 451 050/REV A/03-035

3.7 Results storage and displayTo store the current displayed result press the STO key. The instrument display will momentarilyshow STORED on the seco

Página 9 - 451 050/REV A/03-036

Section 4Maintenance4.1 GeneralThe Model 4510 is designed to give optimum performance with minimum maintenance. It is onlynecessary to keep the extern

Página 10 - 451 050/REV A/03-03

15Section 6Interfacing6.1 AnalogueAll units are provided with 2 x 4mm sockets, marked as ANALOG OUT, on the rear panel. An analogueoutput voltage of 1

Página 11

16Interfacing (continued)The RS232 communications parameters on the computer or printer need to be set to match those ofthe Model 4510, as detailed be

Página 12

SAFETYPlease read this information carefully prior to installing or using this equipment.1. The unit described in this manual is designed to be operat

Página 13

6.4 PrintingA 32 column serial printer (037 701) is available for use with the Model 4510.Connect the printer via the cable supplied with the printer

Página 14 - 11 451 050/REV A/03-03

18451 050/REV A/03-036.4.1 Example Printout4510 Header printout 4510 Results printout

Página 15

Section 7Troubleshooting7.1TroubleshootingFault Possible Cause ActionNo display Check power supply Check that correct 9V ac power supplyis connected a

Página 16

EC Declaration of ConformityJENWAY Model 4510 Conductivity/TDS Meter complies with the following European Standards:EN 50081-1:1992 Electromagnetic co

Página 17

Model 4510Conductivity/TDS MeterOperating ManualContentsSection 1 IntroductionInstrument Description 1.1Instrument Specification 1.2Section 2 Installa

Página 18

Section 1Introduction1.1 Instrument DescriptionA general purpose Conductivity/TDS/temperature bench meter for use in laboratory situations. Theinstrum

Página 19

451 050/REV A/03-032Section 2Installation2.1 UnpackingRemove the Model 4510 from the packaging and ensure the following items are included:1. Model 45

Página 20 - 17 451 050/REV A/03-03

451 050/REV A/03-0332.3 Displays1. Standard selection – indicates which type of standard is being used. Also indicates when acalibration is being perf

Página 21 - 6.4.1 Example Printout

2.4 Keypad451 050/REV A/03-0341. ESC used to switch the instrument on and to place into standby mode (only if powersupply lead remains connected to th

Página 22 - 19 451 050/REV A/03-03

2.5 Inputs/Outputs451 050/REV A/03-0351. Conductivity Socket 7 pin DIN socket which allows the conductivity cell to be connected.2. Analog Out 2 x 4mm

Página 23

Section 3Operation3.1 Good Practice Guidelines1. For greatest accuracy ensure no particulate matter is suspended in the solution under test. Ifnecessa