Model 4510Conductivity/TDS MeterOperating Manual451 050/REV A/03-03
3.2 Instrument Set-UpThe following section details the set-up modes available to the user.These are: Cell constant, Coefficient, Reference Temperature
3.2.2 Temperature Coefficient set upThe display will show:The secondary display will show COEFF and will then scroll TEMPERATURE COEFFICIENT after10 s
3.2.4 TDS EC RatioThe display will show:The secondary display will show TDS EC and will then scroll TDS EC RATIO after 10 seconds.Symbol will fl
3.2.6 Baud RateFor full details relating to serial protocol refer to Section 6 of this manual.The display will show:The secondary display will show BA
3.5 Calibration with Standard SolutionCalibration of the unit and cell with standard solutions will only be necessary if:1) The cell constant is unkno
3.6 Sample MeasurementGeneralConductivity is a temperature dependent measurement. All substances have a conductivitycoefficient which varies from 1% p
3.7 Results storage and displayTo store the current displayed result press the STO key. The instrument display will momentarilyshow STORED on the seco
Section 4Maintenance4.1 GeneralThe Model 4510 is designed to give optimum performance with minimum maintenance. It is onlynecessary to keep the extern
15Section 6Interfacing6.1 AnalogueAll units are provided with 2 x 4mm sockets, marked as ANALOG OUT, on the rear panel. An analogueoutput voltage of 1
16Interfacing (continued)The RS232 communications parameters on the computer or printer need to be set to match those ofthe Model 4510, as detailed be
SAFETYPlease read this information carefully prior to installing or using this equipment.1. The unit described in this manual is designed to be operat
6.4 PrintingA 32 column serial printer (037 701) is available for use with the Model 4510.Connect the printer via the cable supplied with the printer
18451 050/REV A/03-036.4.1 Example Printout4510 Header printout 4510 Results printout
Section 7Troubleshooting7.1TroubleshootingFault Possible Cause ActionNo display Check power supply Check that correct 9V ac power supplyis connected a
EC Declaration of ConformityJENWAY Model 4510 Conductivity/TDS Meter complies with the following European Standards:EN 50081-1:1992 Electromagnetic co
Model 4510Conductivity/TDS MeterOperating ManualContentsSection 1 IntroductionInstrument Description 1.1Instrument Specification 1.2Section 2 Installa
Section 1Introduction1.1 Instrument DescriptionA general purpose Conductivity/TDS/temperature bench meter for use in laboratory situations. Theinstrum
451 050/REV A/03-032Section 2Installation2.1 UnpackingRemove the Model 4510 from the packaging and ensure the following items are included:1. Model 45
451 050/REV A/03-0332.3 Displays1. Standard selection – indicates which type of standard is being used. Also indicates when acalibration is being perf
2.4 Keypad451 050/REV A/03-0341. ESC used to switch the instrument on and to place into standby mode (only if powersupply lead remains connected to th
2.5 Inputs/Outputs451 050/REV A/03-0351. Conductivity Socket 7 pin DIN socket which allows the conductivity cell to be connected.2. Analog Out 2 x 4mm
Section 3Operation3.1 Good Practice Guidelines1. For greatest accuracy ensure no particulate matter is suspended in the solution under test. Ifnecessa